Why are PIC transfers non-engaging (rotational)?

All PIC transfers are non-engaging by design. Learn why PIC system is made for screw-retained restorations, when the connection orientation matters, and what to do when the screw access emerges buccally.

In short: all PIC transfers are non-engaging by design. PIC system is made for screw-retained multi-unit restorations, which use non-engaging connections so the framework seats passively.

PIC transfers are non-engaging (rotational)

All PIC transfers for direct-to-implant use are non-engaging, whatever anti-rotational feature the implant connection has (external or internal hex, tri-lobe, octagon, indexed conical connection, etc.). They capture the position and angulation of each implant, but not the orientation of the connection. PIC transfers for transepithelial abutments (MUA, SRA, etc.) are non-engaging too, because those abutments are already rotational. We do not offer engaging PIC transfers.

Why?

  • The PIC system is designed for screw-retained restorations, especially multi-unit and full-arch frameworks.
  • Splinted, screw-retained restorations on several implants use non-engaging connections. An engaging connection on multiple implants would stop the framework from seating passively.
  • The PIC file gives the exact position and angulation of every implant, which is everything a passive-fit, screw-retained framework needs. Accuracy and fit are not affected.

When would the connection orientation matter?

Only for custom abutments in cemented workflows. PIC system is not intended for that. For it, use an engaging scan body with an intraoral scanner, or a conventional impression.

Why screw-retained instead of cemented?

  • Global passive fit without cement. In multi-unit restorations, custom abutments plus a cement space have traditionally been used to absorb small misfits between the framework and the implants. The PIC system gives a passive fit across the whole framework, so there is no misfit to compensate and no need to cement.
  • Easier maintenance and hygiene. Cemented restorations tend to collect more calculus and organic debris, and any residual cement is a known risk factor for peri-implant inflammation. Screw-retained restorations avoid both.
  • Retrievability. A screw-retained restoration comes off just by unscrewing it. Removing a cemented restoration usually needs a crown remover, which delivers repeated impacts to the restoration and the implants.

What if the screw access emerges buccally?

Use an angulated screw channel (angled screw / ASC) from the implant or component manufacturer. The restoration stays screw-retained.